Custom RF data logger

Applications: Custom RF Data Logger is suitable for Lyophilizers.

Custom design real time data logger (wireless) for monitoring temperature during the freeze-drying process. Highly efficient (quick validation time), highly reliable (real time data), impressive accuracy (±0.1°C).

 

Description

Specifications:

Sensor element

4 wires PT-100
– flexible PTFE / SS 480 mm (19”)
– customized up to 1000 mm (39”)
Operating Range
– standard -90 °C to 140 °C
– extended -90 °C to 350 °C (with thermal shield)
Accuracy ±0.1°C from -90 to 140°C
Memory capacity 32,000 data points
Sampling rate 1 sec to 24 hours
Internal clock drift 4 sec / 24 h @ 23°C
Logger update rate 5 sec to 24 h – user configurable
Mechanical
– material Peek 1000 – FDA compliant
– enclosure IP-68
– weight 115g (4.06 oz) and 125g (4.41 oz)
– small (L/h/Ø) 62/70/36 mm (2.44/2.76/1.42”)
– large (L/l/Ø) 62/88/45 mm (2.44/3.46/1.77”)
Power 2 x Lithium batteries
0.35 g/each Lithium content
Classified UN3091 under IATA DGR section 3.9.2.6
Compliant with ICAO/IATA packing instructions 970
User replaceable
Battery lifetime Up to 6 months, based on usage
Wireless range 121 m (400 feet) line of sight
Radio frequency 2.4 Ghz ISM band
Calibrations Factory: Traceable NIST/COFRAC – ITS-90 coefficients stored in the internal memory
User: Close loop calibration, using XpertLog® software
Certification FMC: EN 60601-1-2 / 2007
RF: EN 300440-2 v1.3.1 (2009-03)
Health Safety: EN 60950-1 / 2006

Optional Accessories:

  • Pliers (Tools to open loggers for battery replacement)
  • Bracket (Mounting bracket for standard and RF data loggers)
  • Access Point (Access Point for real time wireless data loggers)
  • Repeater (Repeater to extend the coverage area of the Access Point)
  • Activation Point (Activates GMP RF and LMS wireless dataloggers)

Additional information

Specifications

Sensor element4 wires PT-100– flexiblePTFE / SS 480 mm (19”)– customizedup to 1000 mm (39”)Operating Range– standard-90 °C to 140 °C– extended-90 °C to 350 °C (with thermal shield)Accuracy±0.1°C from -90 to 140°CMemory capacity32,000 data pointsSampling rate1 sec to 24 hoursInternal clock drift4 sec / 24 h @ 23°CLogger update rate5 sec to 24 h – user configurableMechanical– materialPeek 1000 – FDA compliant– enclosureIP-68– weight115g (4.06 oz) and 125g (4.41 oz)– small (L/h/Ø)62/70/36 mm (2.44/2.76/1.42”)– large (L/l/Ø)62/88/45 mm (2.44/3.46/1.77”)Power2 x Lithium batteries 0.35 g/each Lithium content Classified UN3091 under IATA DGR section 3.9.2.6 Compliant with ICAO/IATA packing instructions 970 User replaceableBattery lifetimeUp to 6 months, based on usageWireless range121 m (400 feet) line of sightRadio frequency2.4 Ghz ISM bandCalibrationsFactory: User:Traceable NIST/COFRAC – ITS-90 coefficients stored in the internal memory Close loop calibration, using XpertLog® softwareCertificationFMC: RF: Health Safety:EN 60601-1-2 / 2007 EN 300440-2 v1.3.1 (2009-03) EN 60950-1 / 2006

Accessories

Optional Accessories: Pliers (Tools to open loggers for battery replacement) Bracket (Mounting bracket for standard and RF data loggers) Access Point (Access Point for real time wireless data loggers) Repeater (Repeater to extend the coverage area of the Access Point) Activation Point (Activates GMP RF and LMS wireless dataloggers)