Description
Sensor element | Absolute pressure | |
4 wires PT-100 | ||
– flexible | – 480mm (19″) PTFE/SS | |
– customized | – up to 5000mm (195″) | |
Operating Range | 0.01 to 5 BarA | |
-40 °C to 140 °C | ||
Accuracy | ±10 mbar ≤ 3 BarA @ 130°C | |
±30 mbar > 3 BarA @ 130°C | ||
±20 mbar ≤ 3 BarA @ 23°C | ||
±50 mbar > 3 BarA @ 23°C | ||
±0.1°C from -40 °C to 140 °C | ||
Memory capacity | 16,000 data points pressure / 16,000 data points temperature | |
Sampling rate | 1 sec to 24 hours | |
Internal clock drift | 4 sec / 24 h @ 23 ° | |
Mechanical | ||
– material | Peek 1000 – FDA compliant | |
– enclosure | IP-68 | |
– weight | 77g (2.72 oz) | |
– diameter | 40 mm (1.57″) | |
– height | 96 mm (3.78″) | |
Power | 1 Lithium battery | |
0.35 g Lithium content | ||
Classified UN3091 under IATA DGR section 3.9.2.6 | ||
Compliant with ICAO/IATA packing instructions 970 | ||
User replaceable | ||
Battery lifetime | Up to 6 months, based on usage | |
Docking station | USB connection | |
4, 8, 12 and 20 ports available | ||
programs and reads unlimited numbers of loggers | ||
Calibrations | Factory: | Traceable NIST/COFRAC – ITS-90 coefficients stored in the internal memory |
User: | close loop calibration, using XpertLog® software | |
Certification | FMC: | EN 60601-1-2 / 2007 |
Health Safety: | EN 60950-1 / 2006 |
Optional Accessories:
- Pliers (Tools to open loggers for battery replacement.)
- Bracket (Mounting bracket for standard and RF data loggers)
- Shelf Mapping Cap (Temperature transfer adapter from the shelf to the logger’s sensor during freeze dryers mappings)
- GMP Pipe (Adapter for fixing data logger’s sensor inside pipes)
- Docking Station (Docking station for GMP and GDP data loggers)