Kaye ValProbe Temperature Data Logger (Bendable) (1 Month Rental)

The Kaye ValProbe is a wireless stainless steel data logger system that collects temperature in environments that are hostile to traditional sensor wires or that are difficult to access. The Kaye ValProbe system is designed for accurate, convenient, and reliable process measurements for pharmaceutical, medical device, and food processing applications. These applications include steam, dry heat and EtO sterilizers, ovens, incubators, stability and temperature chambers, autoclave and warehouses.

Description

The Kaye ValProbe is a wireless stainless steel data logger system that collects temperature in environments that are hostile to traditional sensor wires or that are difficult to access. The Kaye ValProbe system is designed for accurate, convenient, and reliable process measurements for pharmaceutical, medical device, and food processing applications. These applications include steam, dry heat and EtO sterilizers, ovens, incubators, stability and temperature chambers, autoclave and warehouses.

Data from multiple wireless Loggers is combined in a single encrypted file from which reports are generated. These reports are designed for implementation of 21 CFR Part 11 requirements for electronic records. Users can customize the reports by defining process cycles and specifying cycle-based calculations.

The Kaye ValProbe Loggers accurately measure temperature from -45 ᵒC to 140 ᵒC using a precision Platinum RTD element. The RTD element is housed in the probe tip and ensures the following accuracy:

Temperature Accuracy
0 ᵒC to 140 ᵒC ± 0.1 ᵒC
0 ᵒC to -45 ᵒC ± 0.25 ᵒC

Data Collection:

Each Kaye ValProbe Logger collects and stores up to 10,000 data samples in EE PROM memory at a user-specified sample rate. The Loggers can be programmed to start data collection immediately, at a specified time or temperature, or at a specified time and temperature. Table below shows the time ranges available for programming Logger sample rates, start times, and stop times:

Logger Function Minimum Maximum
Sample Rate 1 second 12 hours
Delay Start 0 second 3 days
Delay Stop 1 minute 10 days
Data points Up to 10,000

Dimensions:

Inches
Diameter 1.25
Height 1.75
Length 12